General
Any EMS DolorClast® devices must be maintained/repaired by an authorised EMS Repair Centre.
Your entire unit control unit, handle, compressor requires service 2 years from purchase and then yearly after.
Annual service is completed on-site in 2 hours by our EMS Service Partner.
Where no EMS technician is available in your area, your entire unit will need to come into our workshop for annual service.
- Power Cables
- Control Unit
- Compressor
- Hoses
- Handpiece (No need to send in all applicator heads, one is enough)
Just send in your handle with an applicator head. You don't need to send all of your applicator heads, as you can still use them on the loan handpiece you receive.
If on-site annual service is unavailable our service partners U-Tech Medical will organise a courier to pick up your unit and have it delivered directly to our workshop
Please follow the instructions from U-Tech to help guide you.
Do not use any alcohol-based cleaning solution or warm water to wipe down the unit.
DolorClast® devices should only use EMS contact gel
Compared to other contact gels, EMS’ gel is high in viscosity and doesn’t dry as sticky. This ensures that the gel does not enter the chamber of your handle and causes it to seize.
Control Unit
Ensure that your unit is connected to power supply and that there is no tension on the electrical cord.
Make sure to check the horizonal bar where you select the amount of shocks you want to deliver. Most cases 1 or 2 LEDs will stay solid and not flash like the rest of the LEDs.
If all lights are flashing, the power supply is causing the issues and should be sent in for repair.
Handpiece connection issue.
Unplug the handpiece and make sure it is connected properly with the red dot matching up on the handpiece connector and control unit.
Compressed air issue.
Ensure you check the compressor is plugged in and turned on.
Check air hose connector at rear of unit, make sure this is securely fitted.
If still audible, organise a service/repair. Most likely, the internal valve has become faulty.
Compressor
Check that it is plugged in.
Check that the switch on top of the compressor is switched on (dial turned to the right).
There is a plastic bottle that catches all the condensation and gunk from the compressor lines. Just pull the bottle forward and unscrew from the lid.
This can be poured straight down the sink as it is 95% water.
Dependant on weather and changes of season. Best practice is to monitor it weekly and never let it reach more than half full.
Letting it get too high can cause the fluid to be sucked back into the system and cause multiple compressor and control unit issues.
Handle
Ensure that the applicator head has one o-ring facing the front and one o-ring facing the back. The front one is larger, and the backward facing is smaller.
Helps project the handpiece parts from stress fractures.
Provides a barrier to stop EMS gel entering cavities that can cause projectile issues.
The spanner indicates that your handpiece requires a refurbishment.
- Radial: Requires refurbishment every 1 million shocks
- Power+: Requires refurbishment every 600,000 shocks
- EVO Blue: Requires refurbishment every 5 million shocks (but you need to install a maintenance kit every 1 million shocks)
Projectile is getting stuck and needs to be changed with new revision kit if an EVO handle.
Record the sound and add as a sound bite to service site.
For EVO you can remove the revision kit and submerge in water.
Please send in your handle for repair/refurbishment/display repair.
Display is faulty and needs to come in for repair.